Basic Tutorial
Advanced Tutorial
Download the latest installer for the ASL 5000® Lung Solution LLEAP plug-in (version 2.9) as well as all current product documentation below.
Please uninstall old ASL 5000® Plug-ins prior to updating LLEAP or the ASL 5000® plug-in to avoid functionality issues
For customer with LLEAP version 7.3 and below, please use ASL plug-in version 2.7.12 found at the bottom of the page
- LLEAP Plug-in (version 2.9.9.7)
- ASL 5000® Lung Solution User’s Manual
- Quick Connect Guide for SimMan Family
- Quick Set-up Guide for SimMan Family
- Quick Connect Guide for SimBaby
- Quick Set-up Guide for SimBaby
- Quick Connect Guide for SimMan ALS
- Quick Set-up Guide for SimMan ALS
- Quick Connect Guide for Nursing Anne Simulator
- Quick Set-up Guide for Nursing Anne Simulator
- IngMar Medical pre-configured scenarios for SimMan 3G (Adult)
🩺 Release Notes – Lung Solution LLEAP Plug-in
What’s New:Â
-
n/a
Improved:
-
Fixed issues with connections to an incorrect ASL 5000 device ID
In process and known bugs:
-
Sometimes upon the start of a simulation, the selected condition does not apply until the user clicks the “apply” button a second time
-
In certain situations, the Live Data PIP reading may display incorrectly depending on the waveform shape
Coming soon:
-
n/a
Functionality notes:
-
Peak Inspiratory Pressure (PIP) reading from ASL Monitor window may display differently from what is being shown on the ventilator when using Volume Control ventilation. Peak Pressure on ASL Monitor waveform should match closely to PIP from ventilator.
-
The user may experience minor waveform artifact in the ASL Monitor waveforms when using NIV such as CPAP or BIPAP. This artifact is likely attributed to the ASL 5000 sensing turbulent flow and pressure around the NIV mask.
-
Readings from the ASL 5000 with a manikin and ventilator are attempted to be kept within a 10% +/- accuracy range for values such as Flow, Pressure, and Volume. However, since the readings are being taken at two opposite ends of the system there may be instances where the values fall outside this 10% +/- range. These instances can include the use of Lung Models where high amounts of air trapping are present, instances with patient/ventilator asynchrony, the use of ventilators for which a circuit compensation test has not been performed, as well as leaks in the system such as ventilator circuit leak, ETT leak, internal tubing disconnection, or a stuck lung valve. Volume conversion settings may also have to be changed in LLEAP to match the specified settings of the ventilator in use (As Measured vs. BTPS, for example).
