New Manikin Interface for ASL

Product features

Options

Specifications

How it works

Aerosol testing

FAQ

Support

Download brochure

Request pricing

 

 


 

Combined active and passive breathing simulator with digitally controlled piston for simulation of compliances and resistances via a pressure feedback loop. (Part No. 31 00 100)

Modes of Operation

Parameter file scripts

Interactive (parameters changed directly by the user)

Time-varying parameters
(parameter changes adjusted breath by breath in a pre-scripted fashion)

External interactive
(parameter change algorithms provided by user-written software)

Pause mode (user-definable idling state)

Lung Model Types

Passive (1 or 2 compartment)

Spontaneous

Combined (with adjustable response to ventilator breaths)

Flow waveform generator

Hardware
Total volume

3.1 L

Tidal volumes

2 to 2700 cm “FRC” setting adjustable (200 – 1500 cm3)

Spontaneous
Breath rate

0 to 150 / min (infant Vt)

Peak flow

180 L / min ± 10% (t90Flow < 50 ms)

Resistance settings

3 to 500 cm H2O / L / s, (linear and parabolic resistor types)

Compliance settings

0.5 to 1000 mL / cm H2O

Breath profiles

user-defined pressure data files (breath profiles)
or parameter-selected standard patterns 

Leak settings

possible through optional SBLVM Simulator Bypass and Leak Valve Module

Small signal

better than 15 Hz bandwidth        (10 cm3 response to HF ventilation)

Airway pressure

uncertainty < than 1% fso measurement

Barometric pressure

uncertainty < 1% (1 kPa)

Gas temperature

uncertainty < 0.5 Co (20 – 45 Co)

Input/Output
Input


2 channels analog input 0-10 V 16 bit solution
(can be used for auxiliary pressure and flow signals)

Output

5 V TTL-signal for breath phase recognition, (may be used for synchronizing external data acquisition system with the ASL 5000 simulator)

Software   
Interface to PC Host


via RS 232 at 57.6 kBaud

Host

LabVIEWTM based software environment
With an integrated set of applications for:

-modeling-simulation 
-data analysis

Servo update rate

2048 Hz

Electrical

Universal input,  90…240 V AC

Dimensions

425 x 219 x 315 mm

Weight

10 kg (approx.) 22 lbs.

Patent

Protected under US patent 5975748

 

Simulator Bypass and Leak Valve Module (SBLVM)
External accessory for setting 3 levels of leaks. Also switches between simulator and an external breathing bag when no simulation is running to avoid alarming ventilator.
(Part No. 31 00 200)

Leak settings

approximately 4, 9, 15 L/min at 20 cm H2O, leak orifices exchangeable

Dimensions

4.8 x 4.8 x 4.3 inches / 123 x 123 x 108 mm

Weight

3.5 lbs. (1.6 kg)

Fast Oxygen Measurement (FOM)
For comprehensive ventilator evaluation. Delivers reading almost instantaneously from a highly accurate paramagnetic oxygen sensor in sidestream configuration. Integrated sample pump and barometric pressure sensor for pressure correction.
(Part No. 31 00 300)

Measurement range

0 to 100%

Uncertainty

less than + 0.5% O2

Response time

<350 msec (t10/90 21 to 100% O2)

Cylinder Temperature Controller (CTC)
For simulations requiring realistic physiological conditions. Foil heater with auto-tune PID controller, wall temperature probe, and independent safety thermostat (120 or 240 V). (Part No. 31 00 400)

Wall temperature
Setting

ambient to 45o C

Carrying Case

Hardshell transport case with foam insert, retractable handle and wheels. (Part No. 31 00 500)

Dimensions

25 x 20 x 14 inches / (63 x 51 x 36 cm)

Color:

black

Auxiliary Gas Exchange Cylinder (AGEC)

Bag-in-bottle accessory for use of ASL 5000 with aggressive chemicals. 

Designed to work with bags and bellows < 5” OD. (Part No. 31 00 600)

Volume

approximately 3L

Mobile Cart
Ergonomic design holds laptop or a digital projector and 17 or 18" LCD flat panel display. Four twin wheel swivel castors for maximum stability.
(Part No. 31 00 700)
 



IngMar Medical, Ltd. • Pittsburgh, PA USA • Tel: (412) 441-8228 • Toll free (800) 583-9910 •