Passive lung simulator with two independent
compartments for simulating pediatric and adult
patients. Three 2.5" pressure gauges, range -10
to 100 cmH2O with peak pressure
indicator
(drag pointer). Not intended for ventilator
calibration or performance verification. (Part No. 20 00
300)
|
|
Volume |
2 x 1.0 L nominal |
|
Compliances |
18 to 80 mL/cmH2O
overall range |
|
Resistances |
switch-selectable
15, 25, 50 cmH2O/L/s |
|
Leakage
|
switch-selectable
10, 16, 25 L/min for both ET-tube and
lung leak at 40 cmH2O |
|
Dimensions |
11.5 x 9.8 x 6.5 inches
(291 x 248 x 165 mm) |
|
Weight |
14.5 lbs (7 kg) to 17.5 lbs. (8 kg)
depending on installed options |
|
Color (enclosure) |
blue |
|
 |
|
|
Spontaneous Breathing (SB 2000) Module
Servo motor drive and lever system integrated
into lung model for simulating spontaneous
breathing. Controlled by a handheld unit.
(Part No. 27 00 000)
|
|
Parameters
Breath rate
|
1-60/min
|
|
Tidal volume |
50-500 ml |
|
Peak flow rate |
60 L/min |
|
Renders a breath-to-breath view of inspiratory effort
Tidal volume offset to accommodate ventilator
PEEP settings |
|
Electrical supply |
External power supply included |
|
Pulmonary Mechanics Graphics (PMG 3000) Module
Pressure and flow monitoring system integrated
into lung model. Connects via RS-232 to a
personal computer. Includes software for data
acquisition and display.
(Part No. 26 00 000)
|
|
Hardware
Flow sensors
|
2 fixed orifice sensors, one located
in "main airway", one in right lung
branch |
|
Flow range |
1.8 to 180 L/min
|
|
Accuracy
|
± 3% |
|
Minute volume |
2 to 180 L/min |
|
Tidal volume |
100 to 3000 mL
|
|
Airway pressure |
-120 to 120 cmH2O |
|
Electrical |
120/230 V wall mount charger,
UL/VDE listed |
|
Graphic
Displays
Airway flow *aw
(t)
Airway pressure Paw
(t)
Lung flow *RL (t)
Lung pressure PRL
(t)
Pressure /volume
loops
Flow/volume loops
|
Parameters
Minute Ventilation
Tidal Volume (insp., exp.)
Peak expiratory flow
Airway Resistance
Dynamic compliance
Peak insp. pressure
Mean airway pressure
PEEP
AutoPEEP
Work of Breathing
Respiratory Rate
I:E-ratio |